SunTech 247™ BP + Vitals Device

SunTech 247™ BP + Vitals Device

SunTech 247

Overview

Automated BP devices have become routine for blood pressure measurement. The 247™ BP + Vitals device offers SunTech®’s clinical-grade blood pressure technology in a package designed to fit easily into clinical workflows for measuring and collecting patient vital sign information. With its modular design, start with blood pressure and add other parameters such as temperature and SpO2 when needed. Additionally, the SunTech 247 comes in a variety of mounting options so you can use it wherever, whenever and however you like.

A 247 Diagnostic Station equipped with the optional Bluetooth™ capability can send blood pressure, temperature, and oxygen saturation data via a Bluetooth™ wireless EMR/EHR interface.

Patient population Adult and pediatric patients (age 3 and above).
Method of measurement Oscillometric
Initial inflation pressure 160mmHg +/- 20mmHg
Blood pressure range (mmHg)  60 < Systolic BP < 270, 30 < Diastolic BP < 170
Blood pressure accuracy Measurements determined with this device are equivalent to those obtained by a trained observer using the cuff/stethoscope auscultation method, within the limits prescribed by the American National Standard, Electronic or automated sphygmomanometers.
Blood pressure determination time 35-40 seconds typical for Adult cuff
Heart rate range 30-200 bpm +/- 2% or +/- 3 bpm, whichever is greater
Temperature range 86°F (30.0°C) – 109.4°F (43.0°C)
Temperature accuracy +/- 0.2°F (+/-0.1°C)
Functional oxygen saturation range 40-100%
Functional oxygen saturation accuracy 70-100% +/- 2 digits
(Note: because pulse oximeter equipment measurements are statistically distributed, only about two-thirds of pulse oximeter equipment measurements can be expected to fall within ± 2 digits of the value measured by a CO-oximeter.)
Operating conditions 10°C (50°F) to 40°C (104°F) Less than 90% RH
Storage conditions -20°C (-4°F) to 55°C (131°F) Less than 90% RH
Power External power supply, p/n 19-0013-00 or equivalent approved by SunTech. Input: 100-240V AC @ 47-63 Hz. Output: 6V DC @ 15W.
External power supply, p/n 19-0016-00 or equivalent approved by SunTech. Input: 100-240V AC @ 50-60 Hz. Output: 9V DC@ 18W.
Calibration Check once per 2 years for BP and Temperature
Safety systems Independent hardware over-pressure circuit and redundant software overpressure algorithm to limit cuff pressure to less than 330 mmHg. Independent hardware timing circuit and redundant software timer algorithm to limit the duration of a blood pressure cycle to less than 180 seconds.
Dimensions Length = 5.5 inches, Height = 11.5 inches, Width = 3.8 inches; Length =14.0 cm, Height = 29.2 cm, Width = 9.7 cm
Standards UL60601-1, CAN/CSA C22.2 601-1
IEC 60601-1, IEC 60601-1-2 (EMC), IEC 60601-1-4, ISO 9919, AAMI SP10:2002, ASTM E 1112, EN 12470-3
Meets “Non-invasive Sphygmomanometers — General Requirements & Supplementary Requirements for Electro-Mechanical BP Measuring Systems”, EN 1060-1, EN 1060-3
Classification Protection against electric shock: Class II (for non-battery version), Internally Powered Equipment (for battery version); Applied parts: Type BF; Mode of operation: Continuous

The manufacturer’s web page may be found here.